Comparison of high frequency, in-situ water quality analysers and sensors with conventional water sample collection and laboratory analyses: phosphorus and nitrogen species

Steven J. Granger, Juan A. Qunicke, Paul Harris, Adrian L. Collins, and Martin S. Blackwell

Abstract. The long-term collection of water samples for water quality analysis with high precision laboratory instrumentation is routine in monitoring programmes however, such sampling is labour intensive, expensive, and therefore undertaken at a low temporal frequency. Advances in environmental monitoring technology however, mean that it is now possible to collect high temporal frequency measurements for a wide range of water quality parameters without the need for the physical collection of a sample. The downside to this approach is that the data can be subject to more noise, due to environmental and instrument variables. This raises the question of whether high frequency, lower precision data are better than low frequency, higher precision data. This study reports the findings of an investigation of agricultural land drainage comparing measurements of total phosphorus (TP), total reactive phosphorus (TRP), ammonium (NH4-N) and total oxidised inorganic nitrogen (NOx-N) collected using both equipment in situ and concurrent water samples analysed in the laboratory. Results show that both in situ PHOSPHAX TP and NITRATAX NOx-N instruments can provide comparable data to that measured using samples analysed in the laboratory; however, at high discharge and low NOx-N concentrations, the NITRATAX can under report the concentration. In contrast, PHOSPHAX TRP and YSI sonde NH4-N data were both found to be incomparable to the laboratory data. This was because concentrations of both parameters were well below the instruments accurately determinable level, and because the laboratory data at low concentrations were noisy.

Steven J. Granger, Juan A. Qunicke, Paul Harris, Adrian L. Collins, and Martin S. Blackwell
 
Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
 
Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Steven J. Granger, Juan A. Qunicke, Paul Harris, Adrian L. Collins, and Martin S. Blackwell
Steven J. Granger, Juan A. Qunicke, Paul Harris, Adrian L. Collins, and Martin S. Blackwell

Viewed

Total article views: 2,067 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,508 502 57 2,067 69 78
  • HTML: 1,508
  • PDF: 502
  • XML: 57
  • Total: 2,067
  • BibTeX: 69
  • EndNote: 78
Views and downloads (calculated since 26 Jan 2018)
Cumulative views and downloads (calculated since 26 Jan 2018)

Viewed (geographical distribution)

Total article views: 1,842 (including HTML, PDF, and XML) Thereof 1,784 with geography defined and 58 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 
Latest update: 25 Apr 2024
Download
Short summary
This manuscript looks at the measurement of water quality parameters through the collection of physical samples which are then analysed in an analytical laboratory compared to data generated by equipment placed in the environment. The advantage with this is that large volumes of data are collected at relatively low cost, however the instruments are not subject to the same scientific rigour as in the laboratory. A comparison of the two allows an assessment to be made of the two approaches.